Technical Specification:

Mattress Dimensions: 40cmX64cm

Width: 105 cm

Depth: 64 cm

Characteristics

Power: 220 VAC (±10%), 50-60 Hz

Tredelenburg: ±12°

Air Curtain

Skin Temperature Monitoring for Twin Babies

Temperature and system alarms

Skin Temperature Alarm Limit: ±0.5, ±1.0, ±1.5°C

Skin Temperature Alarm Limit: ±0.5, ±1.0, ±1.5°C

Air Temperature Alarm Limit: ±0.5, ±1.0, ±1.5°C

Skin probe disconnection/ Failure

Air Circulation Failure

Power Supply Failure

System failure

Air sensor failure

Servo-Active Humidity

Measurement range: 0 to 100%

Reservoir capacity : 1 liter

Refilling time: 24h @ 85%

System accuracy : ±1.8%

Display resolution: 1% RH

Adjustment Range: 30 to 95% RH

Humidity alarms Limits: ±5, ±10, ±15 (%)

Fill alarm limit

Sensor Failure Alarm

Temperature Control

Air mode control range: 20°C-39°C

Skin mode control range: 34°C-38°C

Temperature resolution: 0.1°C

Measurement Accuracy(Skin/Air Temp.): 0.2°C

Temperature increase: <35 minutes